Natural Convection Incubator BGI2F3 (BIGP-6203)

Heating Incubator, Heating Natural Convection Incubator
  • Sea, Air, Door to Door Shipping
  • 1 Year Warranty
  • US & European Standards



Natural Convection Incubator Touch screen 124L

Specifications

ModelBGI2F3
TypeTouch screen
Capacity (L)124L
Temp RangeRT+5 - 80°C
Stability±0.2°C
Uniformity (at 37°C)±0.8°C
Display Resolution0.1°C
Max number of shelves2(10)
Max load per shelf20Kg
NW79Kg
Timing Range1 - 5999mins
Interior Dimension WxDxH (mm)520x480x490
Exterior Dimension WxDxH (mm)810x588x755
Electrical RequirementAC220V/50Hz
Power Consumption550W

Description

Heating Incubator - natural convection:
The heating incubator provides stable and controllable temperature environment through precise temperature control. It adopts natural convection mode, stable airflow and small temperature fluctuation. It is widely used in bacterial culture, fermentation, breeding, constant temperature culture and other tests.

Heating Incubator - forced convection:
The heating incubator provides stable and controllable temperature environment through precise temperature control. It adopts forced convection mode, full airflow exchange, good temperature uniformity and small fluctuation. It is widely used in bacterial culture, fermentation, breeding, constant temperature culture and other tests.

Features

Intelligent:
It provides two options: colorful touch screen controller and LCD controller. It is easy to operate. (Professional type LCD controller)
With preset power on, standby and shutdown functions.
Parameters such as multi-stage temperature, circulating wind speed, time and heating rate can be set and programmed at the same time, simplifying the complicated test process and realizing automatic control and operation.

Specialization:
Tested and manufactured using the international DlN-12880-2007 standard.
Provide professional temperature control, high temperature control accuracy and small temperature uniformity error.
The sample can be viewed through the internal glass door without affecting the temperature inside the box (option)
Stackable to save lab space.
Flexible shelf design for efficient use of interior space.

Safety:
Protection of equipment: The second set of temperature limit alarm system conforms to international standards. When the heating is out of control or exceeds the maximum limit temperature, the heating is automatically cut off, and the sound and light alarm reminds the operator. Ensure that the equipment operates safely without accidents.
Protection of key components: The key electrical components are equipped with over-current, over-temperature, overload and other safety protection to prevent accidental equipment.
Protection of the sample: When the temperature inside the box is higher or lower than the set temperature, the alarm will start to cut off the heater, and the sound and light will remind the operator to protect the sample from normal test without accident.
Protection for users: The cabinet and door are specially insulated to make the surface temperature of the cabinet low, ensuring the operator's safety and no accidents.
Provide fault information: When the device fails, the display will display fault information to ensure that the fault information is clear at a glance.

Humanized design:
Adopting direct heating system, the heating speed is fast and the control precision is high.
PID control technology, high control precision and small temperature fluctuation.
Available in Chinese and English menus to meet different language needs, with ° °C/°F conversion.
The height of the shelf is adjustable to meet different cultivation needs.
Professional stacking: Tool design, the machine can be stacked, saving lab space and improving the use efficiency (same specification).

Convenient data processing:
The touch screen type standard with a USB Interface, which can record the change status of the temperature parameters. The LCD screen is option for USB Interface.
It can be equipped with RS232 data Interface, which can realize remote control of the machine through software (option).
RS-232 and USB can be selected one of them.

Safety:
Protection on Instruments: Comply International standard secondary temp limiting alarm system. When the heating is out of control or exceeds the maximum limit temperature, the heating is automatically cut off with sound and light alarm, ensure operating is safe without any accident.
Protection on key components: Key components are equipped with over-current, over-temperature, overload and other safety protection to prevent accident.
Protection on samples: When the temperature inside the box is higher or lower than the set temperature, the alarm will start to cut off the heater, with sound and light alarm to remind the operator to protect the sample from normal test without accident.
Protection on operator: The cabinet and door are specially insulated to make the surface temperature of the cabinet low, ensuring the operator's safety and no accidents.
Breakdown message provided: When the instrument breakdown, the message will show on the screen to help operator clearly check.


Colorful intelligent touch screen:
Safety: Protection on Instruments: Comply International standard secondary temp limiting alarm system. When the heating is
Advanced Color Intelligent Touch Screen Control
1. 4.3 inch touch screen, intelligent touch control, real-time display of parameters, simple and convenient operation.
2. Quick setting of temperature, time and other parameters can be carried out.
3. It has the function of program setting. It can be programmed in 7 segments, 63 steps, 9 steps and 1-99 hours and 59 minutes per segment.
4. With touch screen automatic locking function to prevent non-related personnel from disoperation.

Professional LCD Controller 1. Intelligent program control with high brightness LCD screen, button operation and real-time
Professional LCD Controller
1. Intelligent program control with high brightness LCD screen, button operation and real-time display of parameters.
2. Quick setting of temperature, time and other parameters can be carried out.
3. It has the function of program setting. It can be programmed in 7segments, 63 steps, 9 steps and 1-99 hours and 59 minutes per segment.

Precise temperature control:
Precise temperature control: The uniformity of the temperature inside the incubator chamber is ≤±10 degree , so that all
The uniformity of the temperature inside the incubator chamber is ≤±10 degree , so that all samples are heated evenly.
Note: The stability and uniformity are the result under steady state.

Program control mode, linear temperature rise control. Natural Convection Incubator Touch screen 124L
Program control mode, linear temperature rise control.

Temperature fluctuation inside the incubator chamber is ±0.2 degree, which ensures the stability of the experiment. Natural
Temperature fluctuation inside the incubator chamber is ±0.2 degree, which ensures the stability of the experiment.

Program control mode, Step heating control. Professional duct design: With a unique duct design, the use of large impeller
Program control mode, Step heating control.
Program control mode, Step heating control. Professional duct design: With a unique duct design, the use of large impeller
Professional duct design:
With a unique duct design, the use of large impeller fans, fast temperature rise and fall, stable temperature control, good temperature uniformity.
The circulating fan has adjustable speed, middle and low grades, so as to avoid the volatilization of samples due to excessive air flow during the test.
Famous brand fan, high efficiency, low noise, energy saving and environmental protection.

Stainless steel inner liner: 304 stainless steel, mirror polishing, strong anti-corruption ability. Large arc angle design,
Stainless steel inner liner:
304 stainless steel, mirror polishing, strong anti-corruption ability.
Large arc angle design, no dead angle, easy cleaning and maintenance.

Ultraviolet sterilization system (option) Ultraviolet lamp is located on the upper wall of the box body. It can disinfect
Ultraviolet sterilization system (option)
Ultraviolet lamp is located on the upper wall of the box body. It can disinfect the box body regularly and kill the plankton bacteria in the circulating air of the box body effectively so as to prevent the pollution during cell culture.

No overturning system: Wire shelves with no overturning system to always operate safely. Natural Convection Incubator Touch
No overturning system:
Wire shelves with no overturning system to always operate safely.

Temperature test hole: External detection sensor can real-time detect the temperature inside the box through the test hole
Temperature test hole:
External detection sensor can real-time detect the temperature inside the box through the test hole to ensure that the temperature inside the box is accurate and reliable.

Temperature limit alarm system: When the temperature control system is out of control and the temperature exceeds the limit
Temperature limit alarm system:
When the temperature control system is out of control and the temperature exceeds the limit
temperature, the heating is automatically interrupted to ensure the safety of testing.

Sample temperature sensor (optional): Optional sample temperature sensor, real-time detection of sample temperature, so that
Sample temperature sensor (optional):
Optional sample temperature sensor, real-time detection of sample temperature, so that the temperature is more real and reliable.
Double Convection Model:
Fan power can be adjusted from 0% (equivalent to natural convection) - 100% four gears, which can meet the training requirements of different airflow.

Photos

BIGP-6203BIGP-6203BIGP-6203

Operating Manual

Download


1.Warranty
2.Contents of package
3.Installation requirements and safety tips
    3.1 Installation requirements
    3.2 Instruction for Safety
4.Product introduction
    4.1 Function introduction
    4.2 Specifications

1.Warranty

Thank you for purchasing a Biolab instrument. In normal use conditions, the instrument is guaranteed for a period of 12 months from the date of purchase.

The warranty is valid only if the product is original. It does not apply to any product or parts of it that have been damaged due to incorrect installation, improper connections, improper use, accident or abnormal conditions of operation.

The manufacturer declines all responsibility for damage caused by failure to follow instructions, lack of maintenance and any unauthorized modification

2.Contents of package

The instrument is delivered complete with the following parts:

- Incubator (main unit)

- 2 stainless steel wire shelves

- 4 brackets for shelves

- Power supply cable

- 2 core of fuse

- User manual

3.Installation requirements and safety tips

3.1 Installation requirements

The incubator should be installed in follow conditions:

1. Dry, clean and stable work table with a flat horizontal surface.

2. Respect minimum spaces around instrument 80 cm.

3. Room temperature between 41°F(5°C)and 104 °F(40°C), and relative humidity maximum of 85%.

4. Power supply socket with earth connection.

5. Power feed between 220-240V - 50-60Hz.

3.2 Instruction for Safety BIGP-6203

3.2.1 Danger!

The improper use of this unit may cause property damage and/or personnel injury.

1. The product must be properly electrically grounded (The Hot line or the Neutral line should not be the grounded connection adhere to the product's requirement before using).

2. Please ensure the voltage and frequency of the power supply are compatible with the incubator power requirements prior to use. The fluctuations of the supply voltage shall not exceed 10% of the nominal supply voltage.

3. This unit must use the included electrical cord with a dedicated electrical circuit with a confirmed electrical ground connection.

4. The power switch MUST be in the "OFF" position when power is connected or disconnected from unit.

5. Do not arbitrarily lengthen or shorten the power supply connection wire. Do not modify the power cord in any way.

6. It is prohibited to put in flammable, explosive, volatile, and corrosive substances for drying and baking.

7. Do not touch the chamber door, the chamber body or the surrounding surface when the set temperature is over 176°F(80°C)!

8. Do not put hands or objects into the air inlet or air outlet.

9. The unit should have routine inspections and should be serviced by a qualified service technician when needed.

3.2.2 Warning!

1. Please use the socket connecting with the ground connection to prevent electric shock. If the socket does not have the ground connection, the earth wire must be installed by the qualified electrician. Be sure not to conduct the ground connection through the gas pipe, water pipe, telephone line or lightning rod! This kind of ground connection may cause electric shock due to the incomplete loops.

2. 304 stainless steel material is not acid resistance, so please pay attention to the corrosion prevention measures. Never place corrosive materials inside the unit to prevent damage

3. Do not tension the power supply cord when plug in.

4. The power cord must be removed from receptacle when any of the following occur:

When replacing the fuse;

When the product is waiting for overhaul due to faults;

When the product goes out of service for a long time;

When the product is being moved;

3.2.3 Caution!

1. The incubator should be located on a strong solid surface.

2. Take care when opening and closing the door to prevent damage to delicate internal components.

4.Product introduction

4.1 Function introduction

Tags:, Description:, Attributes-End, Thank you for purchasing a Biolab instrument. In normal use conditions, the instrument


Natural Convection Incubator BIGP-6203

4.2 Specifications

Natural Convection Incubator


Specification

Requirement

Our Model

Compliance

Deviation

Notes

Type of System

Fully Automated, Random-Access Clinical Chemistry Analyzer with STAT Functionality

Fully Automated, Random-Access with STAT Functionality

Yes

-

The requirement matches the quoted model.

Design

Bench Top / Floor Top, Open System

Bench Top design

Yes

-

bench-top model

Reagent System

Integrated with on-board reagentcooling

90 refrigerated reagent positions

Yes

-

Reagent cooling is specified as 2-8°C in both systems.

Reagent Capacity

32 reagent positions for 10 ml, 25ml, 50 ml bottles

90 Reagent positions

Yes

-

Our model has a higher capacity (90 positions).

Reagent Volume per Test

250 μl per test

10-350 μl per test

Yes

-

Within acceptable range.

Sample System

Outer segment with 50 barcode readable positions, inner segment with 12 positions

60 or 90 sample positions, built-in barcode system available

Yes

-

Meets the requirement with extra capacity in sample positions.

Sample Volume

1-30 μl

1-70 μl

Yes

-

Sample volume exceeds requirement.

Pipetting System

1000 μl syringe, programmable in 1 μl steps

1-70 μl sample syringe, programmable in 0.1 μl steps

Yes

-

The quoted model has more precise pipetting for sample volumes.

Cuvette Wash

Fully automatic, overflow-level detector

Automatic cuvette washing

Yes

-

It has systems provide automatic cuvette washing.

Light Source

Quartz-iodine lamp 12V-20W

12V/30W halogen-tungsten lamp

Yes

-

Light sources are similar.

Wavelength Range

340 - 800 nm

340, 405, 450, 480, 505, 546, 570,600, 630, 700, 750, 800 nm

Yes

-

Matches wavelength range requirements.

Photometric Range

-0.1 to 3.0 Absorbance, Resolution 0.001 Abs

0-3.5 Abs, 0.001 Abs resolution

Yes

-

Both systems are in compliance with photometric range.

Measurement Mode

Kinetic, mono- and bichromatic end-point, two-point measurement

Kinetic, Fixed-time, Rate (Kinetic), 1- and 2-point, non-linear calibration

Yes

-

Model offer a range of measurement modes.

Quality Control

Up to 15 controls, Westgard rules, Levey-Jennings plots

Real-time QC, Westgard multi-rule, Cumulative Sum Check, Twin Plot

Yes

-

Model offer quality control mechanisms.

Quality Standard

ISO Certified, CE

CE, ISO 9001, ISO 13485

Yes

-

Both are compliant with ISO, CE.