Laboratory Vertical Autoclave BEO1E10
Vertical Autoclave, Deodorization- Sea, Air, Door to Door Shipping
- 1 Year Warranty
- US & European Standards

Specifications
Model | BEO1E10 |
Capacity (L) | 100 |
Dimensions (LxWxH) mm | 620x745x1200 |
Chamber dimension (DiaxH) mm | Φ400x700 |
Rated power (W) | 4600 |
Basket size (pcs) | (Φ298x195)x3 |
Power requirement | 220V±10% 32A 50Hz/60Hz |
Temperature display accuracy | 0.1°C |
Working environment | 5°C~40°C Relative humidity <85% |
Chamber material | SUS304 stainless steel (316L as option) |
Sterilization temp. range | 105°C-135°C |
Sterilization time range | 1-6000 min |
Melting temp. range | 60°C-115°C |
Melting time | 1-6000 min |
Warming temp. | 45°C-79°C |
Warming time | 1-9999 min |
Memory function | 60 program saved in system,999 cycles saved by USB |
Appointment timer range | Auto start up appointment)0 ~ 15 days |
Controller | "SMART III" microcomputer controller |
Exhaust speed control | Fully automatic internal discharging with 6 levels choice |
Cooling way | Fast cooling fan |
Pressure gauge range | -0.1-0.5MPa |
Pressure valve set pressure | 0.29Mpa |
Vacuum pump | Pulse: 1-9, Vacuum ≤-80kpa, Leak rate ≤0.13kpa |
Safety device | Lid check, dry protection, over-temp/pressure, electric safety, cooling lock |
Features
Advanced
Colorful LCD screen
LCD screen with English and Chinese choice, using text, data, and graphical methods to dynamically display various statuses, the sterilization process is clear at a glance.

6 level of exhaust speed control
After sterilization, you can choose a different exhaust speed, and you can adjust it freely at any time of exhausting.

Condenser (Option)
Equipped with a professional condenser, the steam generated during the sterilization process is converted into water through the condenser, achieving zero discharge of water vapor/mist.

F0 value calculation
FO value 1s automatic calculated, can print out by micro-printer

5 level access permission
The operation permissions for instruments are divided 1nto five levels, with each level of user
having an independent password and multiple IDs that can be assigned. Each level of user
has limited access to certain usage permissions

Fast cooling fan
To shorten cooling process after sterilization

Auto water fill(Option)
Offer pipe and pump to connect to the direct water supply to achieve automatic water feeding, must be added while production.

Deodorization
A unique odor treatment system that can absorb odors and remove smell generated during the sterilization process, ensuring fresh laboratory air.

Air inlet filter
Filtration 1s higher than 99.99% 1n particles with diameter of 0.2 μm
Safety
1. Nice look
The lid and bench corners are covered with insulation to avoid getting scald
2. Lid closing check
The machine can not start if the lid is not locking well by checking mechanically and electrically.
3. Interlock
Only when the temperature and pressure are safe inside, can the door be opened. If the door is not closed in place, it cannot start working
4. Triple protection for water shortage
The bottom of the sterilization chamber is equipped with three different dry burning protection devices: liquid expansion type, copper temperature sensing type, and electric ion type (water level sensor), to avoid misjudgment caused by a single method.
5. Double protection for over pressure
Mechanical pressure gauge and electric sensing system, once pressure is abnormal, the system will cut power, release the pressure and send out an alarm.
6. Auto troubleshooting
System will monitor the working status, and will cut the power, send out alarm, and report the error information once any abnormal situation is detected.
7. Safety test function
Regular testing of system security and safety protection measures is possible
8. Temperature monitor
Power cut and alarming sent, if there is temperature is too high or temperature rising is abnormal
9. Electrical production
Prevention for over current, short circuit and electricity leakage
10. Saturated steam monitoring
The system automatically monitors the purge out of cold air to ensure the pure steam
Intelligent
1. Calibration port
Offer temp. and pressure interface, together with the 3Q adapter joints, can insert max 15 sensors one time to do the calibration work
2. Saturated steam monitoring
The system automatically monitors the purge out of cold air to ensure the pure steam environment
3. Program memory storage system
Users can set their own program, and save it in the system.
4. Special software for drug packaging materials (option)
The product meets the 2020 Chinese Pharmacopoeia or EP10.2 standards
Hommonization
1. Just right depth
Conforming to ergonomics,the chamber depths of 29/54/80/100L chamber are 378/725/670/780 (mm), making it easy to take and place samples and maintain the instrument
2. Front water tank
Front water tank, easy to plug in and out.
Accessories For Purchase
images | name | description |
![]() | Micro printer | To print temp. time and FO value |
Printing set:printer+digital
pressure sensor:can print temp.and pressure in data and flowchart way

pressure sensor:can print temp.and pressure in data and flowchart way




(3Q)adapter joint