Laboratory Vertical Autoclave BEO1C1
Vertical Autoclave, Basic- Sea, Air, Door to Door Shipping
- 1 Year Warranty
- US & European Standards

Specifications
Model | BEO1C1 |
Capacity(L) | 29 |
Dimension(L*W*H)(mm) | 580x490x960 |
Chamber dimension(Dia*H)(mm) | Φ 320 x 378 |
Power requirements | 220V ± 10% 16A 50HZ / 60HZ |
Rated power(W) | 2600W |
Sus304 basket size(pcs) | (Φ 298 x 260) x 1 |
Chamber material | SUS304 stainless steel |
Sterilization temperature | 105°C~135°C |
Sterilization time range | 1~300min |
Pressure gauge range | -0.1MPa~0.5MPa |
Pulse times | 0~9 times(SE/SR only) |
Exhaust temperature range | 73°C~104°C(SE/SR only,liquid mode) |
Safety valve set pressure | 0.29MPa |
Drying | Drying temperature: 80°C~160°C Drying time: 1~300 min |
Controller | Microcomputer control system |
Safety device | Lid closing checking, over temperature protection system,dry scorch protection, over pressure protection, safety valve, over current/short circuit/electric leakage protection,scald protection, automatic failure-inspection system |
Features
1.Simple and easily understandable control panel to show the sterilization temperature, sterilization time, drying time, press separate keys to set the temperature and time makes the operation easier2. Each sterilization stage is displayed with LED lights, making the sterilization process clear at a glance.
3.Microcomputer intelligent control system ,achieves fully automatic control of the sterilization process.
4. The system automatically monitors the discharge of cold air to ensure a pure steam sterilization environment and the best sterilization effect.
5.Up to 20 error information records can be stored, facilitating instrument management.
Safety

1. Lid closing checking:
Machine can not start if lid is not locking well,the self-induction interlock device can avoid risk of misuse
2. Unique anti-scald cover:
Nice look and avoid causing scald
3. Triple protection for lack of water:
The bottom of the sterilization chamber is equipped with three different dry burning protection devices: liquid expansion type, copper temperature sensing type, and electric ion type (water level sensor), to avoid misjudgment caused by a single method.
4. Double protection for over pressure:
Mechanical pressure gauge and electric sensing system, once pressure is abnormal, the system will cut power, release the pressure and send out an alarm.
5. Auto troubleshooting:
The system will monitor the working status, and will cut the power,send out an alarm, and report the error information once any abnormal situation is detected.
6. Safety test function: Regular testing of system security and safety protection measures is
Possible
7. Real time monitor:
The system will monitor the lid locking, pressure and temperature status on a real time basis to guarantee the safe and successful work of the autoclave.
8. Electrical Protection:
Equipped with a highly sensitive electrical protection system, effectively providing leakage protection, overcurrent and short-circuit protection
9. Note:FD-SA has liquid mode, but can not preset the exhaust temperature for liquid mode.
Accessories For Purchase
images | name | description |
![]() | Micro printer | To print temp.,time and FO value |
Printing set:printer+digital
pressure sensor:can print temp. and
pressure in data and flow chart way


pressure sensor:can print temp. and
pressure in data and flow chart way