Inductively Coupled Plasma Emission Spectrometer solid-state RF power supply, with function of automatch 27.12 MHz ± 0.05%

Specifications

ModelBJN1I1
RF Power technical parameter
- Circuit typesolid-state RF power supply, with function of automatch
- Frequency27.12 MHz ± 0.05%
- Frequency Stability< 0.1 %
- Power Output800 W - 1500 W
- Power Output Stability< 0.3 %
- Escaped RF radiation30 cm away from the instrument, electric field: E < 2V/m
Sampling System Technical Parameter
- Output working coil inner diameter25 mm
- Torque tubeThree concentric, external diameter 20 mm
- Coaxial nebulizerOuter diameter 6 mm
- Double barrel atomizing chamberOuter diameter 34 mm
Gas Flow Controls
- Plasma Argon Flowmeter(100-1000) L/h (1.6-16 L/min)
- Auxiliary Argon Flowmeter(100-100) L/h (0.16-1.66 L/min)
- Carrier Argon Flowmeter(100-100) L/h (0.16-1.66 L/min)
- Pressure Maintaining Valve0 - 0.4 MPa
- Cooling WaterTemperature: 20-25 °C, Rate of Flow >5 L/min, Hydraulic Pressure >0.1 Mpa
Technical index of spectrometer
- GratingMiddle step grating, 52.67 lp/mm, 64 sparkle angle
- Wavelength range160-1000 nm
- Numerical apertureF < 8, ultra-high luminous flux to ensure the detection limit and sensitivity of the instrument
- Resolution< 0.0065 nm @ 200 nm
- AstigmatismEquivalent background concentration of 10000 ppm Ca solution at As 189.042 nm <2 ppm
- Light chamberPrecision constant temperature, ±0.1 °C, Distributed nitrogen purging, normal purging 1.8 L/min, fast purging 3.8 L/min
Testing device technical specifications
- DetectorCID
- Target Size27.6 mm x 27.6 mm, 1024 x 1024 addressing detection units
- Reading modeNon-destructive read (NDRO), full reading (FF) and arbitrary read integral (RAI)

Description

Inductively coupled plasma atomic emission spectroscopy (ICP-AES) is very sensitive technique in emission spectroscopy that measures the mass percentage of the metals in the metal/polymer nanocomposites by exciting its metal atoms/ions by using a plasma and analyzing the emission wavelength of the electromagnetic radiation.

Features

Full spectrum direct reading. Precision constant temperature,35 ± 0.1 °C,Distributed nitrogen purging, normal purging 1.8 L/min, fast purging 3.8 L/min.

Applications

Environmental, Metallurgical, Geological, Petrochemical, Pharmaceutical, Food safety.