Specifications
| Model | BJN1I1 |
| RF Power technical parameter | |
| - Circuit type | solid-state RF power supply, with function of automatch |
| - Frequency | 27.12 MHz ± 0.05% |
| - Frequency Stability | < 0.1 % |
| - Power Output | 800 W - 1500 W |
| - Power Output Stability | < 0.3 % |
| - Escaped RF radiation | 30 cm away from the instrument, electric field: E < 2V/m |
| Sampling System Technical Parameter | |
| - Output working coil inner diameter | 25 mm |
| - Torque tube | Three concentric, external diameter 20 mm |
| - Coaxial nebulizer | Outer diameter 6 mm |
| - Double barrel atomizing chamber | Outer diameter 34 mm |
| Gas Flow Controls | |
| - Plasma Argon Flowmeter | (100-1000) L/h (1.6-16 L/min) |
| - Auxiliary Argon Flowmeter | (100-100) L/h (0.16-1.66 L/min) |
| - Carrier Argon Flowmeter | (100-100) L/h (0.16-1.66 L/min) |
| - Pressure Maintaining Valve | 0 - 0.4 MPa |
| - Cooling Water | Temperature: 20-25 °C, Rate of Flow >5 L/min, Hydraulic Pressure >0.1 Mpa |
| Technical index of spectrometer | |
| - Grating | Middle step grating, 52.67 lp/mm, 64 sparkle angle |
| - Wavelength range | 160-1000 nm |
| - Numerical aperture | F < 8, ultra-high luminous flux to ensure the detection limit and sensitivity of the instrument |
| - Resolution | < 0.0065 nm @ 200 nm |
| - Astigmatism | Equivalent background concentration of 10000 ppm Ca solution at As 189.042 nm <2 ppm |
| - Light chamber | Precision constant temperature, ±0.1 °C, Distributed nitrogen purging, normal purging 1.8 L/min, fast purging 3.8 L/min |
| Testing device technical specifications | |
| - Detector | CID |
| - Target Size | 27.6 mm x 27.6 mm, 1024 x 1024 addressing detection units |
| - Reading mode | Non-destructive read (NDRO), full reading (FF) and arbitrary read integral (RAI) |