CO2 Incubator Water Jacketed 655x656x1030 (mm) 25.8x25.8x40.5 (inch) 544x504x681 (mm) 21.4x19.8x26.8 (inch)

Specifications

ModelBFN1H1
Construction
--Exterior dimensions (WxDxH, mm/inch)655x656x1030 (mm) 25.8x25.8x40.5 (inch)
--Interior dimensions (WxDxH, mm/inch)544x504x681 (mm) 21.4x19.8x26.8 (inch)
--Interior Volume (L/cu.ft)185L / 6.5 cu.ft.
--Water Jacket volume43.5L / 1.54 cu.ft.
--Net Weight110kg / 242 lbs
--Interior MaterialType 304, mirror finish stainless steel
--Exterior MaterialCold-rolled steel, powder coated
--Inner doorOne inner door standard
Temperature
--Heating methodWater Jacket
--Temp. control systemMicroprocessor
--Temp. sensorPT1000
--Temp. range5°C above ambient temperature to 55°C
--Temp. uniformity±0.2°C
--Temp. stability±0.1°C
CO2
--Inlet pressure0.1 MPa
--CO₂ control systemMicroprocessor
--CO₂ sensorThermal conductivity / Infrared
--CO₂ range0 to 20%
--CO₂ stability±0.1%
Humidity
--Humidifying systemHumidity pan
--Humidifying sensorStandard
--Relative humidity≥95%
--DisplayIn 0.1% increments
--Water reservoir volume3L
Shelves
--Shelf dimensions (WxD, mm/inch)466x440 (mm) 18.3x17.3 (inch)
--Shelf constructionType 304, mirror finish, stainless steel
--Standard/Maximum shelves3, 11
Fittings
--Access portStandard
--Air filter0.3μm, Efficiency: 99.998% (for CO₂)
--Remote alarm contactsStandard
De-contaminationHEPA filter system
Rated power430W
Power supply220V/50Hz (standard), 110V/60Hz (optional)
Alarm systemPower interruption * High/low temperature * Deviation of CO₂ * RH * Door ajar * Independent overheat protection
Data outputRS232

Description

Water-jacketed:
The large Size CO2 incubator incorporates a water jacketed system. Because of the heat retention characteristics of water, there is no sudden temperature change in the event of an unexpected power failure. A stable temperature environment is ensured.

Features

HEPA filter:
It applies the long term effectiveness of the HEPA filter to protect your cultures. The filter is very efficient to entrap particulates larger than 0.3um at 99.97%. The HEPA filter system runs continuously and within every 60 seconds, the volume of the entire chamber is disinfected. With help of HEPA filter, the air quality achieves Class 100 clean room levels within 5 minutes following a door opening.
HEPA filter: It applies the long term effectiveness of the HEPA filter to protect your cultures. The filter is very

Airflow system:
The optimized air flow system ensures the temperature and CO2 concentration to be stable and uniform within the chamber.

AUTO-ZERO/AUTO-START:
It combines precise CO2 control with a choice of TC or IR sensors. The microprocessor will automatically "Zero" the incubator(IR type) using room air as a reference every 24 hours. Auto-start function for TC type ensures the sensor's baseline automatically reset without manual
adjustment. These features will maintain an accurate CO2 control without worrying about CO: drift.
AUTO-ZERO/AUTO-START: It combines precise CO2 control with a choice of TC or IR sensors. The microprocessor will
HEPA filter and air flow pattern

Automatic control door heater:
The outer door incorporates a door heater which is interlocked with the surrounding temperature monitoring system. This prevents temperature differences between the chamber and the inner door, thereby preventing condensation.

Humidity display and alarming system:
It is able to create a high humidity environment and the relative humidity (RH) is displayed on the panel, readable in 0.1% increments, including low RH programmable alarm(alerts you of need to add water)
Humidity display and alarming system: It is able to create a high humidity environment and the relative humidity (RH) is

Space Utility:
Stackable design takes up less space. Two or three units can be stacked according to available space and usage.
Space Utility: Stackable design takes up less space. Two or three units can be stacked according to available space and

Automatic gas cylinder switchover system:
This system automatically switches from the primary to secondary gas cylinder when CO2 gas level does not change while an injection valve is open.
Automatic gas cylinder switchover system: This system automatically switches from the primary to secondary gas cylinder when